Polymetron 2586sc Analyseur d'oxygène, compatible Claros, Ethernet IP, 110-240 VAC, sans cordon d'alimentation

Polymetron 2586sc Analyseur d'oxygène, compatible Claros, Ethernet IP, 110-240 VAC, sans cordon d'alimentation
Réf. produit: 2586.99.AA7801
Prix unitaire Nous contacter
Livraison en 2 semaines

Save Time on Design

A single design source and one product platform means you spend less time searching for design files or configuring components.

Easy Adoption

The familiar experience with a modern touchscreen, the ability to use your current Hach sensor, and the same footprint as the 9500 Polymetron, make installation and integration of the SC4500 Controller seamless.

The Connectivity Options You Need

The Controller provides local communication to SCADA or a PLC. From analog, to advanced digital protocols, to wi-fi, cellular or LAN, the SC4500 gives you the flexibility to adapt in a rapidly changing world.

Simplify Maintenance and Operation

Common menu guides reduce variability and provide step-by-step procedures for maintenance and calibration. Standard visual alerts across parameters notify operators when troubleshooting is required. The Hach Polymetron 2586sc has a fast response time of less than 60 seconds.
Unlike traditional amperometric techniques that use two electrodes, the Hach Polymetron 2586sc oxygen scavenger analyser uses a three-electrode design; eliminating voltage drift due to the composition of the water. Self-cleaning electrodes reduce maintenance costs and analyser downtime via PTFE beads that prevent deposits on the electrode surfaces.

Contenu de la boîte

Contrôleur installé dans le panneau, électrode de travail, électrode de référence, contre-électrode, manuel d'utilisation de base

Caractéristiques

Alimentation: Without power cord
Alimentation (tension): 110 - 240 V CA
Certifications électriques: EMC

CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)

Safety
CAN/CSA C22.2 No. 61010-1

cETLus safety mark for:
- General Locations per ANSI/UL 61010-1 & CAN/CSA C22.2. No. 61010-1
Communication (optional): Analog:
Five 0-20 mA or 4-20 mA analog outputs on each analog output module
Up to two analog Input modules (0-20 mA or 4-20 mA). Each input module replaces a digital sensor input.

Digital:
Profibus DPV1 module
Modbus TCP
Profinet IO module
Ethernet IP module
Conditions de stockage: -20 - 70 °C, 0 - 95% relative humidity, non-condensing
Contenu de la boîte: Contrôleur installé dans le panneau, électrode de travail, électrode de référence, contre-électrode, manuel d'utilisation de base
Écoulement: 166 - 250 mL/min (10 - 15 L/h) recommended
Exigences d’échantillon: Sample needs to be free of undissolved matter.
Fonctions relais: Scheduler (Timer), Alarm, Feeder Control, Event Control, Pulse Width Modulation, Frequency Control, and Warning
Humidité relative: 10 - 90% relative humidity (non-condensing)
Intervalle d'entretien: Monthly: Calibration and reagent refill
Langues de la notice: Bulgarian, Chinese (PRC), Croatian, Czech, Danish, Dutch, English, Estonian, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Lithuanian, Polish, Portuguese (Brazil), Portuguese (Portugal), Romanian, Russian, Slovak, Slovenian, Spanish, Swedish, Thai, Turkish
Limite de détection: Drift is negligible; 1 ppb
Logiciel disponible: Cloud (Claros-enabled)
Matériel: Counter-Electrode: Stainless steel; Working Electrode: Platinum
Moyens de communication: Claros Ethernet IP
Niveau d'accès: 2
Nombre de voies: 1
Plage de mesure: 0 - 100 ppb carbohydrazide (also known as ELIMIN-OX); programmable

ELIMIN-OX is a registered trademark of Nalco Chemical Co., Naperville, IL.
  0 - 500 ppb hydrazine; programmable
  0 - 500 ppb hydrazine; programmable
Plage de pression: 0.5 - 6 bar (7.2 - 87 psi) or 12 L/h
Plage de température: 5 - 45 °C at 0 - 95% relative humidity (non-condensing)
Poids: 14.6 kg
Raccordement de ligne de purge: 6 x 8 mm (tubing must not exceed 4 feet and must drain straight down)
Raccordements: 4 x 6 mm stainless steel tubing
Réactif: Sample conditioning with diisopropylamine, ammonia, or diethylamine
Référence: Ag/AgCl electrode with electrolyte KCl 0.1 M
Relai : mode opérationnel: Primary or secondary measurement or timer/scheduler
Répétabilité: ± 2% or 1 ppb (whichever is greater)
Sorties digitales: Four electromechanical SPDT (Form C) contacts, 1200 W, 5 A
Température de l'échantillon: 5 - 45 °C
Temps de réponse: < 60 s