GLI Model C33 Contacting Conductivity Controller

ARTICLE OBSOLETE

GLI Model C33 Contacting Conductivity Controller
Réf. produit: C33A1NN

ARTICLE OBSOLETE

Multiple Measurements

Sensor A and B can be independently selected to measure conductivity (μS/cm or mS/cm), resistivity (MΩ • cm or KΩ • cm), or total dissolved solids (ppm or ppb). The measured values or temperatures for Sensor A or B can be displayed separately or together (four total measurements). Both analog outputs can also be displayed. The Model C33 can provide a calculated Sensor A and B measurement (% rejection, % passage, ratio A/B, ratio B/A, difference A-B, or difference B-A). Front panel red LEDs indicate relay on/off status.

Ultra-pure Water Capability

Model C33 accurately measures conductivity down to 0.056 μS/cm and resistivity up to 18.2 MΩ • cm, both at 25.0°C.

Backlit LCD Readout

The two-line, 16 character backlit LCD provides excellent viewing contrast under all lighting conditions.

Menu-guided Operation

The simple keypad and logical menu structure make this analyzer easy to use. Menu screens guide you through setup, calibration, operation, and test/maintenance functions.

Multiple Language Capability

All screens can be selected for display in English, French, German, or Spanish.

Contenu de la boîte

Controller, User Manual

Caractéristiques

Affichage: Ecran LCD rétroéclairé 2 lignes de 16 caractères
Alarme: Alarm: settings for low alarm point, low alarm point deadband, high alarm point, high alarm point deadband, off delay, and on delay
Alimentation (tension): 180 - 260 V CA
Caractéristique électrique (Hz): 50 - 60 Hz
Compensation température: -20 à 200 °C automatique ou manuelle avec sélection pour compensateur de température (Pt1000 RTD ou Pt100 RTD); % linéaire par courbe °C, tables de compensation de température intégrées Ammonium et eau naturelle ou pas de compensation
  à automatique ou manuelle avec sélection pour compensateur de température (Pt1000 RTD ou Pt100 RTD); % linéaire par courbe °C, tables de compensation de température intégrées Ammonium et eau naturelle ou pas de compensation
Conditions de stockage: -30 °C à 70 °C
Conductivité mesurées: sélectionnable :
0-2,000
0-20,00
0-200,0
0-2000
(µS/cm ou mS/cm)
Consommation d'énergie: 10 VA
Contenu de la boîte: Controller, User Manual
Dérive de température: < 0.03 % of span per °C (zero and span)
Déviation: 0.05 % d'envergure par 24h, non cumulatif
Fonctions relais: Statut : non configurable ; le relai n'est activé que lorsque un diagnostic d'échec existe
Garantie: 24 mois
Hauteur: 104 mm
Humidité relative: 0 à 95 % sans condensation
Interface: RS-232
Longueur de câble max. entre le capteur et le transmetteur: 91 m
Matériel: acier zingué
Measuring range: -20 - 200 °C
Méthode d'étalonnage: GLI DRY-CAL Method:
enter the sensor’s GLI-certified “calibration constant” and temperature “T” factor
1-POINT SAMPLE Method:
enter one sample value derived by laboratory analysis or comparison reading
ZERO Method:
With the dry sensor in air, press keys to initiate automatic system zeroing
Montage: Montage sur panneau
Nombre de voies: 2 sensor A or B’s measured values or temperatures can be separately displayed, or all four measurements can be displayed together on
one screen. Both outputs are displayed together on a screen. Also, a calculated sensor A and B measurement can be displayed
Notes spéciales: les DELs relai A et B indiquent respectivement le statut du relai
  Mesure calculée capteur A et B :
0-100% rejet
0-100% passage
ratio A/B ou B/A
différence A-B ou B-A
Plage de température: -20 - 60 °C at 0 - 95% RH (non-condensing)
Poids: 0,8 kg
Précision: 0,1 % d'envergure
Relai : mode opérationnel: Chaque relai (A et B) peut être conduit par la mesure sélectionnée du capteur A ou du capteur B (conductivité, résistivité ou TDS) ou par la température
Répétabilité: 0.1 % d'envergure ou mieux
Sauvegarde de la mémoire: Tous les paramètres de l'utilisateur sont retenu indéfiniment en mémoire (EEPROM) non volatile
Sortie de courant: 2 x 0/4 à 20 mA chacun avec résolution 0,004 mA (12-bit) et charge max. 600 Ω
Sorties digitales: 2 relais électromécaniques ; contacts SPDT (Forme C) ; U.L. évalué 5A 115/230 V AC, 5A à 30 V DC résistive
TDS mesurées: sélectionnable :
0-9999 ppm
0-9999 ppb